PET/SPECT Preformance Incl. PET Lid

SKU
PH0007
Available

Preformance Phantom

Phantom for NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

€ 1.800,00
Product description
Available

Technical data (can be modified to customer specifications):

  • main cylinder:
    – inside cylinder diameter: 206mm
    – inside cylinder height: 186mm
    – cylinder wall thickness: 7mm
  • cold rods insert:
    – rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm
    – height of rods: 88mm
  • cold spheres:
    – solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8mm
    – height of the center of the spheres from the base plate: 127mm
  • PET Lid with cylindrical samples:
    – refillable thin-walled cylinders, diameters: 8, 12, 16 and 25mm
    – water filled cylinder diameter: 25mm
    – air filled cylinder diameter: 25mm
    – PTFE solid cylinder diameter: 25mm
    – cylinder height: 38mm
  • heavy duty carrying case

Product features:

  • Complies with:
    – NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    – AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    – AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    – ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration