Pro-MRI

SKU
PH0029
Available

Phantom for comprehensive evaluation of critical imaging parameters of magnetic resonance imaging (MRI) in a time efficient manner. The phantom can be used for the measurement of absolute values for calibration purposes. However, its design is optimized for time efficient daily quality assurance too.

€ 1.400,00
Product description
Available

It can be used to measure:

  • - Geometric Distortion
  • - Spatial Resolution
  • - Slice thickness and position
  • - Interslice Gap
  • - T1 and T2 values
  • - Estimate of Image Bandwidth
  • - Low Contrast Detectability
  • - Image Uniformity
  • - Signal-to-Noise Ratio (SNR)
  • - Physical and Electronic Slice Offset
  • - Point of reference
  • - Bandwidth: water-fat shift

Technical data (can be modified to customer specifications):

  • - outside cylinder diameter: 220mm (180mm)
  • - outside cylinder height: 150mm
  • - inside cylinder diameter: 204mm (173mm)
  • - inside cylinder height: 130mm
  • - filled with 10 mmol nickel chloride solution containing
  • - sodium chloride 75 mmol
  • T1 and T2 sample vials
    – 6 cylindrical vials Ø19 x 41mm 
    – refillable from outside
  • – by default filled with 20 mmol nickel chloride and 15 mmol sodium chloride solution
  • resolution insert: 
    – four matrices of holes, diameters: 0.8, 0.9, 1.0 and 1.1 mm
    – spaces between the holes are equal to hole diameters
    – two small containers with water and fat for water-fat shift evaluation
  • slice thickness insert:
    – 180 x 46 x 10mm PMMA plate
    – 1mm wide and 5mm deep counter-descending slits on both sides
    – slits form two ramps descending at 1:10 geometric distortion insert:
    – 10 x 10 array of squares 
    – 148mm on a side
    – 10mm thick
  • low contrast insert:
    – 4 polycarbonate discs 0.05, 0.1, 0.15 and 0.2mm in thickness
    – partial volume contribution of these sheets and filling solution produce contrasts: 1.4, 2.5, 3.6 and 5.1%
    – each disc contains 12 groups of 3 holes arranged in spokes 
    – each spoke has the same diameter
    – diameters range from 7.0 to 1.5mm (0.5mm step)
  • four sets of paired 45° wedges are located on both sides of the phantom. The lower pairs are 30 x 30mm and the upper ones are 40 x 40mm. The distance between intersection points of the lower and upper pairs is 90mm
  • optional 6 removable vials for test samples - replacing the "T1 and T2 sample vials”, includes 6 additional vials (12 total)
  • optional heavy duty carrying case

Product features:

  • Complies with:
    – ACR MRI Accreditation Program
    – IPEM Report 80 “Quality Control in MRI”, 1998
    – AAPM Rep. 28 “Quality Assurance methods & phantoms for MRI”, 1990
    – AAPM Rep. 34 “Acceptance testing of MRI systems”, 1992
    – AAPM Rep. 100 “Acceptance & Quality Assurance procedures for MRI facilities”, 2010
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration