._.

Pro-NDT CR

SKU
PH0028
Available

This Image Quality Indicator is used for testing all important parameters of CR scanner systems according to ASTM E 2445-05, ISO 16371-1 and EN 14784-1 standards. It provides means to measure: spatial resolution, unsharpness, contrast, MTF, laser beam jitter, scanner slipping and shading.

€ 2.100,00
Product description
Available

Technical data (can be modified to customer specifications):

  • - dimensions: 350 x 430 x 19 mm
  • - made of PMMA (Lucite)
  • - T shaped brass target (laser beam jitter, MFT check, blooming / flare)
  • - Pro-NDT DuplexWireIQI test object (basic spatial resolution, unsharpness)
  • - BAM snail made of lead and low absorbent material (central beam alignment)
  • - two converging line pairs resolution patterns (Pro-Res MAM ConeType)
  • - three Ø19x0.3mm (EL, EC, ER) measuring points (shading correction)
  • - cassette positioning locator
  • - homogeneous 0.5mm Al strip (scanning slipping, shading)
  • - two cm/inch rulers (linearity check)
  • - three contrast sensitivity gauges made of Al, Cu and stainless steel with four holes of different depth
  • - calibration / test certificate according to ASTM E 2445-05, EN 14784-1 and ISO 16371-1

Product features:

  • complies with:
    – ASTM E 2445-05
    – ISO 16371-1
    – EN 14784-1
  • CE certified (declaration of conformity according to ISO/IEC 17050-1)
  • the Manual provides guidelines for carrying out each test, results assessment and registration
._.